Hoppa till innehåll
Free shipping on orders over 3000 SEK excl. VAT! Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%
Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%A bore can hold size at one point and still be both tapered and ovalThe temperature was right when the load was loaded, but what happened during transport?With a 0.4-second temperature reading, you can measure accurately before the product has had time to changeCore temperature is what counts: how to document cooling and heating processes without handwritten listsThe cold chain failed overnight, but nobody can prove when or for how longUsing the Same Thermometer for Raw and Cooked Meat Is a Risk That Two Colors RemoveFrom ”it looks like a crack” to a measured and documented dimension on screenAntifreeze protection in the cooling system can look right and still be wrong: how to measure glycol content and freezing point in secondsHow to know that the glycol in your cooling circuit actually protects against frostTemperature and humidity in the warehouse can be monitored without anyone going there to take readingsVolume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%A bore can hold size at one point and still be both tapered and ovalThe temperature was right when the load was loaded, but what happened during transport?With a 0.4-second temperature reading, you can measure accurately before the product has had time to changeCore temperature is what counts: how to document cooling and heating processes without handwritten listsThe cold chain failed overnight, but nobody can prove when or for how longUsing the Same Thermometer for Raw and Cooked Meat Is a Risk That Two Colors RemoveFrom ”it looks like a crack” to a measured and documented dimension on screenAntifreeze protection in the cooling system can look right and still be wrong: how to measure glycol content and freezing point in secondsHow to know that the glycol in your cooling circuit actually protects against frostTemperature and humidity in the warehouse can be monitored without anyone going there to take readings

The Setting Ring — the Hole with a Known Diameter That Determines the Accuracy of Everything You Measure

Published August 22, 2026 at 03:41

Most internal measuring instruments don’t measure diameter. They measure deviation — how much a bore differs from a size the instrument has already been set to. A two-point bore gauge, a Kræplin or Schwenk gauge, a bore gauge with a dial indicator: all of them are first zeroed against a known diameter and then used to read off the difference. That’s what makes them fast and easy to read in production.

But it also pushes the whole question of accuracy one step back. The instrument’s reading can never be better than the size it was zeroed against. And that size is a ring.

LMW setting ring in hardened steel according to DIN 2250 form C, 100 mm

What it is

A setting ring in hardened steel according to DIN 2250 form C, nominal diameter 100 mm. The inner surface is precision-ground, and the size is stated in mm with the deviation specified in micrometers — not as a rounded nominal value, but as what the ring actually measures. An included calibration certificate makes the diameter traceable.

The problem it solves

The workflow is simple and worth printing out, since every step matters:

  1. Wipe the ring’s inner surface clean. A fingerprint or a chip is several micrometers thick — more than the entire tolerance you’re working with.
  2. Insert the measuring head until the two-point contact is felt.
  3. Zero the instrument against the ring’s documented diameter — not against the rounded nominal figure, but against the actual one.
  4. Remove the head and measure the workpiece. Now the instrument shows the deviation from a known quantity.

The fact that the deviation is stated in micrometers on the ring is the detail that separates a standard from a manufactured ring. No one grinds a ring to exactly 100.000 mm — you grind it as close as you can get, measure what it turned out to be, and mark that on it. A ring marked ”100 mm, +2 µm” is therefore a reference of 100.002 mm, and that’s the figure the instrument should be zeroed against. Use 100.000 and you’ve built a two-micrometer systematic error into every measurement made afterward — on every part, all day.

The other thing that makes the ring a standard is dimensional stability. Hardened steel that has been ground and aged holds its size year after year, which is the whole point: the reference must not move while the instruments it calibrates are swapped out. That’s also why the same temperature rule applies as for all length metrology — the ring and the instrument must be at the same temperature when zeroing is done, otherwise you’re zeroing against a size that doesn’t exist.

And the ring itself must be checked. The recommendation is periodic verification, typically at twelve-month intervals, in step with the renewal of the calibration certificate. A worn or damaged ring is especially treacherous, because the error it introduces isn’t random but systematic — every measurement is wrong in the same direction, and no internal spot check will reveal it.

Three typical use cases

  • Zeroing in production: daily setup of two-point bore gauges and internal measuring instruments before running a batch.
  • Verification of bore gauges: periodic check that the instrument is still reading correctly, in a calibration lab or in-house.
  • Traceability: as the documented link between the workshop’s measurements and an external calibration.

Why it pays off

This is one of the few products whose value lies entirely in what it makes possible for others. A ring never measures a single part on its own — but every part measured with an instrument zeroed against it inherits its accuracy. Scale that up: a ring costing a few thousand kronor determines the correctness of thousands of measurements a year.

And the financial risk lies in the opposite direction. An error in the reference is systematic, which means it isn’t caught by re-measuring, isn’t caught by having someone else measure, and isn’t caught by comparing parts against each other — they’ll all be equally wrong. It’s only discovered at assembly, at the customer, or at the next external calibration — and then the question covers the entire period since the last one. Keeping the reference clean, temperature-stable, and checked is the cheapest insurance a metrology department can buy.

Read more about the LMW setting ring 100 mm, DIN 2250 form C →